EasyManua.ls Logo

Anritsu MS2840A - [:SENSe]:SWEep:EGATe:LENGth <time>; [:SENSe]:SWEep:EGATe:LENGth?

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 SCPI Device Message Details
2-300
[:SENSe]:SWEep:EGATe:LENGth <time>
Gate Length
Function
This command sets the time length of the gate.
Command
[:SENSe]:SWEep:EGATe:LENGth <time>
Parameter
<time>
Time length of the gate
Range 50 µs to 1 s
Resolution 20 ns
Suffix code NS, US, MS, S
S is used when omitted.
Example of Use
To set the gate time length to 20 ms.
SWE:EGAT:LENG 20MS
[:SENSe]:SWEep:EGATe:LENGth?
Gate Length Query
Function
This command queries the time length of the gate.
Query
[:SENSe]:SWEep:EGATe:LENGth?
Response
<time>
Parameter
<time>
Time length of the gate
Range 50 µs to 1 s
Resolution 20 ns
Suffix code None. Value is returned in s units.
Example of Use
To query the gate time length.
SWE:EGAT:LENG?
> 0.02000000