EasyManua.ls Logo

Anritsu MS2840A - :TRIGger[:SEQuence]:SOURce?

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.7 Sweep/Trigger/Gate
2-273
2
SCPI Device Message Details
:TRIGger[:SEQuence]:SOURce?
Trigger Source Query
Function
This command queries the trigger source.
Query
:TRIGger[:SEQuence]:SOURce?
Response
<source>
Parameter
<source>
Trigger signal source
[MS269xA]
EXT
External input (External)
IMM
Free run
WIF
Wide bandwidth IF detection (Wide IF Video)
VID
Video detection (Video)
SG
SG Marker
BBIF
Baseband Interface (BBIF)
[MS2830A], [MS2840A]
EXT
External input (External)
IMM
Free run
WIF
Wide bandwidth IF detection (Wide IF Video)
VID
Video detection (Video)
SG
SG Marker
FRAM
Frame period trigger
Details
[MS269xA]
SG marker trigger can be selected only when the Option 020/120 Vector
Signal Generator is installed.
BBIF trigger cannot be selected only when the Option 040/140 Baseband
Interface Unit is not installed or the software package is Ver.6.00.00 or
later.
[MS2830A], [MS2840A]
SG marker trigger can be selected only when the Option 020/120/021/121
Vector Signal Generator is installed.
Example of Use
To query the trigger signal source.
TRIG:SOUR?
> VID

Table of Contents

Other manuals for Anritsu MS2840A