Chapter 2 SCPI Device Message Detail
2-836
[:SENSe]:TOI:ZSPan:BANDwidth|BWIDth[:RESolution]:AUTO ON|OFF|1|0
TOI Zero Span Measurement Resolution Bandwidth Auto/Manual
Function
This command turns on and off the automatic setting of the resolution
bandwidth for the Zero Span Measurement in the TOI measurement.
Command
[:SENSe]:TOI:ZSPan:BANDwidth|BWIDth[:RESolution]:AUTO
<switch>
Parameter
<switch>
Automatic setting of the resolution bandwidth
for the Zero Span Measurement in the TOI
measurement
ON|1
Turns on the automatic setting of the resolution
bandwidth for the Zero Span Measurement in
the TOI measurement. (Default)
OFF|0
Turns off the automatic setting of the resolution
bandwidth for the Zero Span Measurement in
the TOI measurement.
Example of Use
To automatically set the resolution bandwidth for the Zero Span
Measurement in the TOI measurement.
TOI:ZSP:BAND:AUTO ON