EasyManua.ls Logo

Anritsu MS2840A - :CALCulate:MARKer:PEAK:THReshold <ampl>

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.5 Signal Search
2-173
2
SCPI Device Message Details
:CALCulate:MARKer:PEAK:THReshold <ampl>
Peak Search Threshold Level
Function
This command sets the threshold when a peak point is detected.
Command
:CALCulate:MARKer:PEAK:THReshold <ampl>
Parameter
<ampl>
Threshold when searching for the peak point
Range Full width of Y axis
Resolution 0.01 dB (When Scale Unit settings are
dB-system.)
0.01 pV (When the setting of Scale Unit is V.)
0.01 yW (When the setting of Scale Unit is W.)
Suffix code
DBM,DM
dBm
DBMV
dBmV
DBUV
dBµV
DBUVE
dBµV (emf)
DBUVM
dBµV/m
V
V
MV
mV
UV
µV
W
W
MW
mW
UW
µW
NW
nW
PW
pW
FW
fW
Follows the setting of Scale Unit, when omitted.
V is used when it is Linear Scale.
Default Y-axis center
Details
This command is not available during Spurious Emission measurement
and when Displayed Segment Mode is Auto.
Not available during Spectrum Emission Mask measurement.
Example of Use
To set the threshold when searching for the peak point to
10 dBm.
CALC:MARK:PEAK:THR 10DBM