Chapter
Native Device Message Details
4-174
Details
This command is not available in the following cases:
During the Spurious Emission measurement and when Displayed
Segment Mode is set to Auto.
During the Spectrum Emission Mask measurement.
Example of Use
To move the zone marker center to 500 points from the screen left edge.
MKZ 500
MKZ?
> 500
MKZF/MKZF?
Zone Marker Frequency (Time)
Function
This command moves the zone marker center to the specified frequency
(time).
Command
MKZF freq,marker
MKZF time,marker
Query
MKZF? marker
Response
freq
Returns a value without a suffix code, in Hz units, and with a
resolution of 0.01 Hz.
time
Returns a value without a suffix code, in
µ
s units, and with a
resolution of 0.01
µ
s.
Parameter
freq
Zone marker center frequency
Range
[MS269xA]
100 MHz to 6.0 GHz (MS2690A)
100 MHz to 13.5 GHz (MS2691A)
100 MHz to 26.6 GHz (MS2692A)
[MS2830A]
100 MHz to 3.7 GHz (Option 040)
100 MHz to 6.1 GHz (Option 041)