2.3 RBW/VBW
2-79
2
SCPI Device Message Details
Details
In MS269xA, MS2830A, MS2840A RBW Mode: Normal
The resolution bandwidth Auto setting is switched off when the
resolution bandwidth (RBW) is changed.
For MS2830A, MS2840A RBW 20 MHz or greater is only available for
MS2830A-005/105/007/009/109, MS2840A-005/105/009/109.
For RBW 31.25 MHz, it is not a Gauss filter but a flat-top
characteristic filter. This command can be used only during 0 span.
RBW 1 Hz, 3 Hz, and 10 Hz cannot be set when Auto Sweep Type
Rules is Swept Only. This command cannot be used during 0 span
because Auto Sweep Type Rules is fixed to Swept Only.
This command is disabled during the Spectrum Emission Mask
measurement.
This command is disabled during the Spurious Emission measurement.
In MS2830A, MS2840A RBW Mode: CISPR
The resolution bandwidth Auto setting is switched off when the
resolution bandwidth (RBW) is changed.
RBW 1 MHz cannot be set when Detector is Quasi-Peak.
Not available when the Measure function is set to On.
Not available when the Gate View function is set to On.
Example of Use
To set the RBW to 3 kHz.
BAND 3KHZ
Related command
This command has the same function as the following commands.
[:SENSe]:ACPower:BANDwidth[:RESolution]
[:SENSe]:CHPower:BANDwidth[:RESolution]
[:SENSe]:OBWidth:BANDwidth[:RESolution]
[:SENSe]:BPOWer|:TXPower:BANDwidth[:RESolution]
:CALCulate:BANDwidth|:BWIDth[:RESolution]
:CALCulate:ACPower:BANDwidth[:RESolution]
:CALCulate:CHPower:BANDwidth[:RESolution]
:CALCulate:OBWidth:BANDwidth[:RESolution]
:CALCulate:BPOWer|:TXPower:BANDwidth[:RESolution]