Chapter
4
Native Device Message Details
4-56
Suffix code NS, US, MS, S
S is used when the suffix code is omitted.
Default 0 s
Details
This command is not available during Spurious Emission measurement
and when Displayed Segment Mode is Auto.
Not available during Spectrum Emission Mask measurement.
Not available when Marker Mode is set other than Delta.
Example of Use
When Frequency Span is 0 Hz, and Active Marker is Maker 2.
To move the center of Marker 1 zone marker to +100 MHz of Marker 2
(reference marker).
CALC:MARK1:REF 2
CALC:MARK:X:DELT 1,100MHZ
CALC:MARK:X:DELT? 1
> 100000000.00
When Frequency Span is 0 Hz, and Active Marker is Maker 2.
To move the center of Marker 1 zone marker to 15 ms of Marker 2
(reference marker).
CALC:MARK1:REF 2
CALC:MARK:X:DELT 1,15MS
CALC:MARK:X:DELT? 1
> 0.015000000