2.4 Marker
2-149
2
SCPI Device Message Details
When scale unit settings are V-system units.
Resolution 0.001 dB
Suffix code None. Value is returned in dBm units.
999.0 is returned when no measurement is
performed or an error occurs.
999.0 is returned when Marker Mode is set to Fixed or Off.
999.0 is returned when Zone Width Type is set
to Spot.
When scale unit settings are W-system units.
Resolution 0.01 yW
Suffix code None. Value is returned in W units.
999.0 is returned when no measurement is
performed or an error
occurs.999.0 is returned when Marker Mode is
set to Fixed or Off.
999.0 is returned when Zone Width Type is set
to Spot.
Details
This command is not available during Spurious Emission measurement
and when Displayed Segment Mode is Auto.
Example of Use
To query the total power and the power density within the range of the
zone marker.
CALC:PMAR:Y?
> -8.000,-50.000