Chapter
Native Device Message Details
4-146
Details
This command executes Transmit intermodulation measurement and
outputs the result based on the specified parameter list file.
Place parameter list files in a folder on the specified drive.
<device>:\Anritsu Corporation\Signal Analyzer\User Data\Batch
If the parameter list file has been changed, the changes must be applied
with the
MMEM:REL:BATC
command.
(cf
. MMEM:REL:BATC
)
Example of Use
To execute Transmit intermodulation measurement using the
MyParam.xls parameter list file.
MEAS:BATC:IM? "MyParam.xml",2GHz,2.005GHz
>
0.0,-72.130,0.0,-72.130,-1.270,-73.400,-0.570,-72.700,-0
.780,-72.910,-1.030,-73.160,-999.0,-999.0,-999.0,-999.0
Format of Parameter list file
Refer to
:MEASure:BATCh:IM?
in “Chapter 2 SCPI Device Message
Details”.