SCPI Device Message Details
When Marker Mode is Delta and Scale Mode is Log
resolution: 0.001 dB
Suffix code None. Value is returned in dB units.
–999.0 is returned at no measurement or error
When Marker Mode is Delta and Scale Mode is Lin
Range 0.0000 to 10000
Resolution 0.0001
Suffix code None, no units,
–999.0 is returned at no measurement or error
Details
This command is not available during Spurious Emission measurement
and when Displayed Segment Mode is Auto.
Example of Use
To query the level at the marker point of Marker 1.
CALC:MARK:Y?
> -73.687
Related command
This command has the same function as the following commands.
:CALCulate:ACPower:MARKer[1]|2|3|4|5|6|7|8|9|10:Y?
:CALCulate:CHPower:MARKer[1]|2|3|4|5|6|7|8|9|10:Y?
:CALCulate:OBWidth:MARKer[1]|2|3|4|5|6|7|8|9|10:Y?
:CALCulate:BPOWer|:TXPower:MARKer[1]|2|3|4|5|6|7|8|9|10:
Y?
:CALCulate:SPURious:MARKer[1]|2|3|4|5|6|7|8|9|10:Y?