2.5 Signal Search
2-179
2
SCPI Device Message Details
UW
µW
NW
nW
PW
pW
FW
fW
Follows the setting of Scale Unit, when omitted.
V is used when it is Linear Scale.
<resolution>
when Scale Mode is Log.
Range 0.001 to 100.000 dB
Suffix code DB
dB is used even when omitted.
when Scale Mode is Lin.
Range 0.01 to 100.00%
Suffix code None
<sort>
AMPLitude
Outputs in the order of level from the highest.
FREQuency
Outputs in the order of level from lowest.(only
frequency domain)
TIME
Outputs in the order of time from the earliest
(only time domain)
When omitted: Outputs in the order of level from highest.
<total_number>
Number of the output peak value
<level_n>
Level value of peak
When marker level display unit are dB-system units.
Resolution 0.01 dB
Suffix code None. Value is returned in the unit specified by
Scale Unit.
999.0 is returned when not measured or an
error occurs.
When marker level display units are V-system units.
Resolution 0.01 pV
Suffix code None. Value is returned in V units.
999.0 is returned when not measured or an
error occurs.
When marker level display units are W-system units.
Resolution 0.01 yW
Suffix code None. Value is returned in W units.
999.0 is returned when not measured or an
error occurs.
<freq_n>
Peak frequency
resolution, 0.01 Hz
Suffix code None. Value is returned in Hz units.
999999999999 is returned when not measured
or an error occurs.