Chapter 2 SCPI Device Message Details
2-242
This command is available only for MS2830A.
The destination trace means the selelcted trace.
Trace-writing mode where Store Ref. is executed is set to View.
This cannot be executed under the following conditions:
• Internal Signal Generator Control function On/Off is Off.
• Normalize function is On.
• The source trace and destination trace are the same.
• The destination trace is the active trace.
• The source trace is blank.
Store Ref. data is initialized under the following conditions:
• The Internal Signal Generator Control function On/Off is changed.
• The selected trace is changed.
• The Trace type writing mode is changed to Write.
• The number of trace point is changed.
• Center/Start/Stop/Span frequency is changed.
• RBW/VBW is changed.
• The wave detection mode is changed.
• The SPA application software is preset.
• The SPA application software is unloaded .
• The SPA application software is recalled.
• Correction On/Off in the Configuration is changed.
When changing the writing method of the trace on which Store Ref. was
executed, use
:TRACe[1]|2|3|4|5|6:TYPE
.
This command toggles between View and Blank of the Ref. Trace Display.
Example of Use
To copy the trace A data to the trace B as Store Ref..
TRAC:COPY TRAC1,TRAC2