Chapter 2 SCPI Device Message Details
2-242
This command is available only for MS2830A.
The destination trace means the selelcted trace.
Trace-writing mode where Store Ref. is executed is set to View.
This cannot be executed under the following conditions:
Internal Signal Generator Control function On/Off is Off.
Normalize function is On.
The source trace and destination trace are the same.
The destination trace is the active trace.
The source trace is blank.
Store Ref. data is initialized under the following conditions:
The Internal Signal Generator Control function On/Off is changed.
The selected trace is changed.
The Trace type writing mode is changed to Write.
The number of trace point is changed.
Center/Start/Stop/Span frequency is changed.
RBW/VBW is changed.
The wave detection mode is changed.
The SPA application software is preset.
The SPA application software is unloaded .
The SPA application software is recalled.
Correction On/Off in the Configuration is changed.
When changing the writing method of the trace on which Store Ref. was
executed, use
:TRACe[1]|2|3|4|5|6:TYPE
.
This command toggles between View and Blank of the Ref. Trace Display.
Example of Use
To copy the trace A data to the trace B as Store Ref..
TRAC:COPY TRAC1,TRAC2