EasyManua.ls Logo

Anritsu MS2840A - Page 809

Anritsu MS2840A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2.14 Spurious Emission Measurement Function
2-799
2
SCPI Device Message Details
0 is returned when judged as Pass, and 1 is
returned when judged as Fail.
<tracedata_m>
n=2 to 21
Returns a comma separated list of the trace data
for the selected segment (where segment
number = n-1).
n=23 to 42
Returns a comma separated list of the trace data
for the selected segment (where segment
number = n-22).
Values have no suffix code, are in dBm units,
and have 0.001 dB resolution.
999.0 is returned when no measurement is
performed.
<judge>
Pass/Fail judgment to the whole segment
0 is returned when judged as Pass, and 1 is
returned when Fail.
999.0 is returned when no measurement is
performed.
Details
This function queries the result of the Spurious Emission measurement
performed lastly. This function does not accompany any sweep, thus this
function is used To query the measurement result in a different type,
when the measurement has already completed.
Use READ command to perform re-measurement with redoing sweep.
A return value of this command varies depending on the compatible
mode.
(cf.
:SYSTem:RESult:MODE
)
Example of Use
To obtain the result of the Spurious Emission measurement (when
Result Mode is A, and when Spurious Emission Result Type is Worst).
FETC:SPUR?
>
0,1,1,135618.00,-64.25,51.25,-13.00,0,2,2,155970.00,-63.
91,50.91,-13.00,0
・・・・