Chapter 2 SCPI Device Message Details
2-684
Parameter
<ref_power>
Reference absolute power
<abs_lower_offset_n>
Peak value of absolute power of lower
Offset-n
<abs_upper_offset_n>
Peak value of absolute power of upper
Offset-n
No suffix code, dBm unit, 0.001 dB
resolution.
“–999.0” is returned when no
measurement is performed.
<rel_lower_offset_n>
Peak value of relative power of lower
Offset-n
<rel_upper_offset_n>
Peak value of relative power of upper
Offset-n
<margin_lower_offset_n>
Minimum value of margin of lower
Offset-n
<margin_upper_offset_n>
Minimum value of margin of upper
Offset-n
No suffix code, dB unit, 0.001 dB
resolution.
“–999.0” is returned when no
measurement is performed.
<freq_lower_offset_n>
Frequency of peak level of lower
Offset-n
<freq_upper_offset_n>
Frequency of peak level of upper
Offset-n
No suffix code, Hz unit, 1 Hz resolution.
“–999999999999” is returned when no
measurement is performed.
<total_judge>
Total judgment result
<lower_offset_n>
Judgment result of lower Offset-n
<upper_offset_n>
Judgment result of upper Offset-n
This command returns 0 when it is
PASS, and returns 1 when it is FAIL.
This command returns“ –999.0 ”when
no measurement is performed.